| Unique ID issued by UMIN | UMIN000063703 |
|---|---|
| Receipt number | R000068409 |
| Scientific Title | Effect of the Time from Syringe Blood Collection to Aliquoting into Vacutainer Tubes on Coagulation Parameters |
| Date of disclosure of the study information | 2026/11/20 |
| Last modified on | 2026/11/19 22:20:27 |
Effects of the Order and Timing of Blood Transfer into Collection Tubes on Blood Coagulation Test Results
Blood Draw Timing and Coagulation Test Study
Effect of the Time from Syringe Blood Collection to Aliquoting into Vacutainer Tubes on Coagulation Parameters
Study on Dispensing Order and Blood Coagulation Tests
| Japan |
healthy adult
| Nursing | Adult |
Others
NO
To determine how time differences arising from the order in which blood is transferred from a syringe into collection tubes affect blood coagulation test results.
Others
the reliability and accuracy of test results
Changes in PT, APTT, fibrinogen, and related coagulation parameters associated with the time elapsed before aliquoting blood after collection.
Observational
| 20 | years-old | <= |
| 50 | years-old | > |
Male
Medication free healthy adult men aged 20 years or older and under 40 years.
1) Individuals with a history of vasovagal reflex during blood collection
2) Individuals who are unable to follow instructions
21
| 1st name | Seira |
| Middle name | |
| Last name | Ichinohe |
Aomori University of Health and Welfare, Japan
Faculty of health science, Department of nursing
030-8505
58-1 Mase Hamadate Aomori, Japan
017-765-2054
s_ichinohe@ms.auhw.ac.jp
| 1st name | Seira |
| Middle name | |
| Last name | Ichinohe |
Aomori University of Health and Welfare, Japan
Faculty of health science, Department of nursing
030-8505
58-1 Mase Hamadate Aomori, Japan
017-765-2054
s_ichinohe@ms.auhw.ac.jp
Aomori University of Health and Welfare, Japan
Ministry of Education, Culture, Sports, Science and Technology (MEXT), Japan
Japanese Governmental office
Research Ethics Committee of Aomori University of Health and Welfare
58-1 Mase Hamadate Aomori,Japan
017-765-2000
auhw_rinri@ms.auhw.ac.jp
NO
青森県立保健大学(青森県)
| 2026 | Year | 11 | Month | 20 | Day |
Published
21
Data from 21 participants were analyzed. One-way repeated-measures ANOVA showed a significant main effect of blood collection tube for PT (p = 0.012, partial eta squared = 0.15) and FIB (p < 0.001, partial eta squared = 0.21). Dunnett's test found no statistically significant differences between the reference tube and any of the four comparison tubes for any of the four outcomes.
| 2026 | Year | 11 | Month | 19 | Day |
The study included 21 participants. Their age was 27.9 +/- 8.5 years, height was 173.8 +/- 5.4 cm, body weight was 70.3 +/- 9.8 kg, and body mass index (BMI) was 23.4 +/- 3.3 kg/m^2. Values are presented as mean +/- standard deviation.
A total of 21 participants were enrolled. Blood collection and measurements were completed for all 21 participants using the reference blood collection tube and each of the four comparison tubes (1-4). All 21 participants were included in the analysis.
No adverse events were observed in any of the 21 participants during blood collection or during the observation period after blood collection.
Measurements are reported as mean +/- standard deviation in the following order: reference blood collection tube, followed by comparison tubes 1, 2, 3, and 4. Mean differences were calculated as the comparison tube value minus the reference tube value. The 95% confidence intervals (CIs) and p values from Dunnett's multiple comparisons test are reported.
1. APTT (seconds)
Measurements: 29.1 +/- 2.7, 29.1 +/- 2.6, 29.3 +/- 2.6, 29.1 +/- 2.6, and 29.1 +/- 2.7.
A one-way repeated-measures ANOVA with the Greenhouse-Geisser correction (correction factor = 0.73) showed no statistically significant main effect of tube type (p = 0.089, partial eta squared = 0.10).
Mean differences from the reference tube (exploratory analysis):
Tube 1: 0.0 (95% CI [-0.2, 0.1]), p = 1.00
Tube 2: 0.1 (95% CI [-0.01, 0.3]), p = 0.10
Tube 3: 0.0 (95% CI [-0.2, 0.2]), p = 1.00
Tube 4: 0.0 (95% CI [-0.2, 0.2]), p = 1.00
2. PT (seconds)
Measurements: 10.4 +/- 0.5, 10.4 +/- 0.5, 10.4 +/- 0.4, 10.4 +/- 0.5, and 10.3 +/- 0.4.
A one-way repeated-measures ANOVA assuming sphericity showed a statistically significant main effect of tube type (p = 0.012, partial eta squared = 0.15).
Mean differences from the reference tube:
Tube 1: 0.0 (95% CI [-0.1, 0.1]), p = 1.00
Tube 2: 0.0 (95% CI [-0.1, 0.1]), p = 1.00
Tube 3: 0.1 (95% CI [-0.03, 0.1]), p = 0.30
Tube 4: -0.1 (95% CI [-0.1, 0.02]), p = 0.20
3. PT-INR
Measurements: 0.99 +/- 0.06, 0.99 +/- 0.05, 1.00 +/- 0.03, 1.00 +/- 0.06, and 0.99 +/- 0.05.
A one-way repeated-measures ANOVA assuming sphericity showed no statistically significant main effect of tube type (p = 0.529, partial eta squared = 0.04).
Mean differences from the reference tube (exploratory analysis):
Tube 1: 0.00 (95% CI [-0.02, 0.02]), p = 1.00
Tube 2: 0.01 (95% CI [-0.01, 0.03]), p = 0.70
Tube 3: 0.01 (95% CI [-0.02, 0.03]), p = 0.90
Tube 4: -0.01 (95% CI [-0.03, 0.02]), p = 0.90
4. FIB (mg/dL)
Measurements: 233.1 +/- 31.0, 233.9 +/- 32.0, 234.6 +/- 31.3, 230.4 +/- 31.0, and 231.10 +/- 31.8.
A one-way repeated-measures ANOVA assuming sphericity showed a statistically significant main effect of tube type (p < 0.001, partial eta squared = 0.21).
Mean differences from the reference tube:
Tube 1: 0.9 (95% CI [-1.9, 3.6]), p = 0.90
Tube 2: 1.6 (95% CI [-1.2, 4.3]), p = 0.40
Tube 3: -2.6 (95% CI [-5.4, 0.1]), p = 0.10
Tube 4: -2.0 (95% CI [-4.7, 0.8]), p = 0.2
For all four outcomes, Dunnett's multiple comparisons test found no statistically significant differences between the reference tube and any of the four comparison tubes.
Undecided.
Completed
| 2025 | Year | 08 | Month | 21 | Day |
| 2025 | Year | 08 | Month | 04 | Day |
| 2025 | Year | 08 | Month | 21 | Day |
| 2026 | Year | 01 | Month | 15 | Day |
| 2026 | Year | 03 | Month | 09 | Day |
| 2026 | Year | 03 | Month | 30 | Day |
| 2026 | Year | 09 | Month | 28 | Day |
From August 5, 2025, to January 8, 2026, 21 healthy adult men were recruited through posters displayed on the campus of Aomori University of Health and Welfare. For each participant, APTT, PT, PT-INR, and FIB were compared across blood samples collected using the reference blood collection tube and four types of comparison tubes
| 2026 | Year | 11 | Month | 19 | Day |
| 2026 | Year | 11 | Month | 19 | Day |
Value
https://center6.umin.ac.jp/cgi-open-bin/ctr_e/ctr_view.cgi?recptno=R000068409