| Unique ID issued by UMIN | UMIN000048251 |
|---|---|
| Receipt number | R000054994 |
| Scientific Title | Verification study for accuracy of non-invasive cardiac output measurement in children |
| Date of disclosure of the study information | 2022/10/01 |
| Last modified on | 2025/01/02 09:46:07 |
Verification study for accuracy of non-invasive cardiac output measurement in children
Verification study for accuracy of esCCO in children
Verification study for accuracy of non-invasive cardiac output measurement in children
Verification study for accuracy of esCCO in children
| Japan |
Pediatric patients under general anesthesia
| Pediatrics | Anesthesiology | Operative medicine |
Others
NO
Compare the accuracy of non-invasive continuous estimated cardiac output measurements (esCCO) with software updates in pediatric patients.
Efficacy
Compare the accuracy of cardiac output measurements between updated esCCO and transthoracic echocardiography
Observational
| Not applicable |
| 20 | years-old | >= |
Male and Female
Scheduled surgery
Obtained informed consent
Congenital heart disease
Not obtained informed consent
70
| 1st name | Kimitoshi |
| Middle name | |
| Last name | Nishiwaki |
Nagoya University Hospital
Department of Anesthesiology
466-8560
65 Tsurumai-cho, Showa-ku, Nagoya, Japan
+81527442340
nishi@med.nagoya-u.ac.jp
| 1st name | Tasuku |
| Middle name | |
| Last name | Fujii |
Nagoya University Hospital
Department of Anesthesiology
466-8560
65 Tsurumai-cho, Showa-ku, Nagoya, Japan
0527442340
plus9@med.nagoya-u.ac.jp
Nagoya University
Self funding
Self funding
Nagoya University Hospital Ethics Committee
65 Tsurumai-cho, Showa-ku, Nagoya, Japan
0527442479
ethics@med.nagoya-u.ac.jp
NO
| 2022 | Year | 10 | Month | 01 | Day |
Published
https://doi.org/10.1213/ane.0000000000007144
80
Cardiac output measurement by esCCO with modified coefficients for pediatric patients showed high accuracy and tracking ability compared with cardiac output measurement by transthoracic echocardiography. In Bland-Altman analysis, the bias (standard deviation), 95% limits of agreement, and percentage error were 0.03(0.28), -0.53 to 0.60, and 18% in the preoperative measurement, and -0.04(0.19), -0.42 to 0.35, and 15% in the postoperative measurement, respectively.
| 2025 | Year | 01 | Month | 02 | Day |
none
Completed
| 2022 | Year | 07 | Month | 01 | Day |
| 2023 | Year | 01 | Month | 31 | Day |
| 2023 | Year | 03 | Month | 01 | Day |
| 2024 | Year | 12 | Month | 31 | Day |
| 2024 | Year | 12 | Month | 31 | Day |
| 2024 | Year | 12 | Month | 31 | Day |
| 2024 | Year | 12 | Month | 31 | Day |
Prospective observational study
| 2022 | Year | 07 | Month | 03 | Day |
| 2025 | Year | 01 | Month | 02 | Day |
Value