Unique ID issued by UMIN | UMIN000037336 |
---|---|
Receipt number | R000039257 |
Scientific Title | R-R interval prediction by a reflection light sensor. |
Date of disclosure of the study information | 2019/08/01 |
Last modified on | 2019/07/11 11:28:40 |
R-R interval prediction by a reflection light sensor.
R-R interval prediction by a reflection light sensor.
R-R interval prediction by a reflection light sensor.
R-R interval prediction by a reflection light sensor.
Japan |
healthy
Cardiology | Adult |
Others
NO
In the absence of arrhythmia, the heart beat is reflected in the pulse in the periphery and can be measured with a transmission type pulse oximeter. However, most of these devices are somewhat wired and limited, so they are not suitable for measurement during work and exercise. It is a reflection type photoelectric pulse wave sensor module and verifies whether the pulse wave interval (PPI) measured by AGVS that can be recorded wirelessly reflects the electrocardiogram RRI.
Efficacy
Not applicable
Peak-to-peak interval at rest and after exercise.
Pulse rate at rest and after exercise.
Observational
20 | years-old | <= |
Not applicable |
Male
Healthy
Those whom can not apply the instruments.
5
1st name | Soh |
Middle name | |
Last name | Nishimoto |
Hyogo College of Medicine
Department of Plastic Surgery
663-8131
1-1 Mukogawa-cho, Nishinomiya, Hyogo
0798-45-6753
nishimot@hyo-med.ac.jp
1st name | Soh |
Middle name | |
Last name | Nishimoto |
Hyogo College of Medicine
Department of Plastic Surgery
663-8131
1-1 Mukogawa-cho, Nishinomiya, Hyogo
0798-45-6753
nishimot@hyo-med.ac.jp
Hyogo College of Medicine
Genial Light Co.Ltd.
Profit organization
The Ethics Review Board of Genial Light
145-1 Tokiwa-cho, Naka-ku, Hamamatsu, Shizuoka
053-456-3630
info@geniallight.co.jp
NO
2019 | Year | 08 | Month | 01 | Day |
https://www.geniallight.co.jp
Published
https://www.geniallight.co.jp
5
In linear regression, PPI=0.99991xRRI+0.00012, R=0.99939 at fingertip. PPI=0.99896xRRI+0.00087, R=0.99933 at forehead.
2019 | Year | 07 | Month | 11 | Day |
4 healthy men.
no problem
none
R-R intervals and EKG intervals
Completed
2018 | Year | 04 | Month | 01 | Day |
2018 | Year | 04 | Month | 10 | Day |
2018 | Year | 04 | Month | 15 | Day |
2018 | Year | 06 | Month | 15 | Day |
In linear regression, PPI=0.99991xRRI+0.00012,R=0.99939 at finger tip. At forehead, PPI=0.99896xRRI+0.00087,R=0.99933.
2019 | Year | 07 | Month | 11 | Day |
2019 | Year | 07 | Month | 11 | Day |
Value
https://center6.umin.ac.jp/cgi-open-bin/ctr_e/ctr_view.cgi?recptno=R000039257
Research Plan | |
---|---|
Registered date | File name |
Research case data specifications | |
---|---|
Registered date | File name |
Research case data | |
---|---|
Registered date | File name |