Scientific Title | The validity of spatiotemporal gait analysis using dual laser range sensors: a cross-sectional study |
Date of disclosure of the study information | 2018/05/24 |
Last modified on | 2018/05/24 19:22:33 |
The validity of spatiotemporal gait analysis using dual laser range sensors: a cross-sectional study
The validity of spatiotemporal gait analysis using dual laser range sensors: a cross-sectional study
The validity of spatiotemporal gait analysis using dual laser range sensors: a cross-sectional study
The validity of spatiotemporal gait analysis using dual laser range sensors: a cross-sectional study
Japan |
Healthy adults
Rehabilitation medicine | Adult |
Others
NO
The aim of present study was to investigate the validity of the proposed method using dual laser range sensors for spatiotemporal gait measurement by comparison to the reference system using a computer-based instrumented walkway system.
Efficacy
Frequency distributions of errors for Foot Contact and Foot Off times
Inter-class correlation coefficient (ICC(2,1) of six spatiotemporal gait parameters (Swing time, Stance time, Double support time, Stride time, Stride length, Step length)
Observational
Not applicable |
Not applicable |
Male and Female
Healthy adults
history of disease in orthopedic, neurophysiologic,and cardiovascular
10
1st name | |
Middle name | |
Last name | Masanobu Iwai |
Kawamura Hospital
Department of Rehabilitation
1-84 Daihannya, Akutami, Gifu, Gifu
058-241-3311
masanobuiwai@gmail.com
1st name | |
Middle name | |
Last name | Masanobu Iwai |
Kawamura Hospital
Department of Rehabilitation
1-84 Daihannya, Akutami, Gifu, Gifu
058-241-3311
kawamurareha@gmail.com
Kawamura Hospital
None
Self funding
Fujita Health University
NO
河村病院(Kawamura Hospital),藤田保健衛生大学(Fujita Health University)
2018 | Year | 05 | Month | 24 | Day |
Unpublished
Completed
2017 | Year | 05 | Month | 29 | Day |
2017 | Year | 08 | Month | 05 | Day |
2017 | Year | 10 | Month | 03 | Day |
2017 | Year | 12 | Month | 21 | Day |
The detection times for Foot Contact and Foot Off estimated by the proposed method did not differ from those measured by the instrumented walkway reference system. In addition, histogram for Foot Contact and Foot Off showed monomodality.
ICC(2,1) of the spatiotemporal parameters excluding swing time (0.15) were moderate (stance time: 0.74; double support time: 0.53; stride time: 0.89; stride length: 0.83; step length: 0.71). The mean errors of all spatiotemporal parameters were small.