| Recruitment status | Completed |
| Unique ID issued by UMIN | UMIN000031318 |
| Receipt No. | R000035752 |
| Official scientific title of the study | Humidification performance of a HME and a heated and humidified high flow (HHHF) for tracheostomized patients with spontaneous breathing |
| Date of disclosure of the study information | 2018/02/15 |
| Last modified on | 2018/10/19 (Ver. 3) |
| Basic information | ||
| Official scientific title of the study | Humidification performance of a HME and a heated and humidified high flow (HHHF) for tracheostomized patients with spontaneous breathing | |
| Title of the study (Brief title) | Humidification performance of a HME and a heated and humidified high flow (HHHF) | |
| Region |
|
|
| Condition | ||||
| Condition | tracheostomized patients with spontaneous breathing | |||
| Classification by specialty |
|
|||
| Classification by malignancy | Others | |||
| Genomic information | NO | |||
| Objectives | |
| Narrative objectives1 | To compare humidification performance of HNME and HHHF |
| Basic objectives2 | Efficacy |
| Basic objectives -Others | |
| Trial characteristics_1 | Confirmatory |
| Trial characteristics_2 | |
| Developmental phase | |
| Assessment | |
| Primary outcomes | Absolute humidity |
| Key secondary outcomes | Tidal volume,end-tidal carbon dioxide (Et CO2), breathing frequency, SpO2, heart rate, and blood pressure. |
| Base | |
| Study type | Interventional |
| Study design | |
| Basic design | Cross-over |
| Randomization | Non-randomized |
| Randomization unit | |
| Blinding | Open -no one is blinded |
| Control | Active |
| Stratification | |
| Dynamic allocation | |
| Institution consideration | |
| Blocking | |
| Concealment | |
| Intervention | ||
| No. of arms | 3 | |
| Purpose of intervention | Treatment | |
| Type of intervention |
|
|
| Interventions/Control_1 | HHHF (heated and humidified high flow) was used for 1 hour
Optiflow, 40L/min |
|
| Interventions/Control_2 | HME (Heat and moisture exchanger) was used for 1 hour
Trach vent+ |
|
| Interventions/Control_3 | HHHF (heated and humidified high flow) was used for 1 hour
Optiflow, 40L/min |
|
| Interventions/Control_4 | ||
| Interventions/Control_5 | ||
| Interventions/Control_6 | ||
| Interventions/Control_7 | ||
| Interventions/Control_8 | ||
| Interventions/Control_9 | ||
| Interventions/Control_10 | ||
| Eligibility | ||||
| Age-lower limit |
|
|||
| Age-upper limit |
|
|||
| Gender | Male and Female | |||
| Key inclusion criteria | We included consecutive adult tracheostomized patients with spontaneous breathing who had been weaned from mechanical ventilation in the ICU. | |||
| Key exclusion criteria | We excluded patients who met the following criteria: 1) aged less than 18 years old, 2) patients who have too much secretions and difficulty in expelling secretions, 3) patients with severe respiratory or hemodynamic instability, 4) who need more than 1L/min of oxygen or fraction of inspired oxygen (FiO2) > 0.25. | |||
| Target sample size | 10 | |||
| Research contact person | |
| Name of lead principal investigator | Nobuto Nakanishi |
| Organization | Tokushima university hospital |
| Division name | Emergency and critical care medicine |
| Address | 2-50-1, Kuramoto, Tokushima 770-8503, Japan |
| TEL | 088-633-9347 |
| nobuto_nakanishi@yahoo.co.jp | |
| Public contact | |
| Name of contact person | Nobuto Nakanishi |
| Organization | Tokushima university hospital |
| Division name | Emergency and critical care medicine |
| Address | 2-50-1, Kuramoto, Tokushima 770-8503, Japan |
| TEL | 088-633-9347 |
| Homepage URL | |
| nobuto_nakanishi@yahoo.co.jp | |
| Sponsor | |
| Institute | Tokushima university hospital |
| Institute | |
| Department | |
| Funding Source | |
| Organization | None |
| Organization | |
| Division | |
| Category of Funding Organization | Other |
| Nationality of Funding Organization | |
| Other related organizations | |
| Co-sponsor | |
| Name of secondary funder(s) | |
| Secondary IDs | |
| Secondary IDs | YES |
| Study ID_1 | 2033 |
| Org. issuing International ID_1 | Tokushima university hospital |
| Study ID_2 | |
| Org. issuing International ID_2 | |
| IND to MHLW | |
| Institutions | |
| Institutions | 徳島大学病院 |
| Other administrative information | |||||||
| Date of disclosure of the study information |
|
||||||
| Progress | |||||||
| Recruitment status | Completed | ||||||
| Date of protocol fixation |
|
||||||
| Anticipated trial start date |
|
||||||
| Last follow-up date |
|
||||||
| Date of closure to data entry |
|
||||||
| Date trial data considered complete |
|
||||||
| Date analysis concluded |
|
||||||
| Related information | |
| URL releasing protocol | https://www.ncbi.nlm.nih.gov/pubmed/30254040 |
| Publication of results | Published |
| URL releasing results | https://www.ncbi.nlm.nih.gov/pubmed/30254040 |
| Results | Ten subjects were enrolled. The temperature with the HME was 29.9 and during heated-and-humidified high-flow use was 35.3 (P<.001). With both the HME and the heated-and-humidified high-flow system, the relative humidity reached 100%; absolute humidity with HME was 30.2 mg/L, and with the heated-and-humidified high-flow system, was 40.3 (P< .001). |