Unique ID issued by UMIN | UMIN000013985 |
---|---|
Receipt number | R000016293 |
Scientific Title | Impact of atherosclerosis on the reliability of LiDCOrapid |
Date of disclosure of the study information | 2014/05/17 |
Last modified on | 2018/01/21 10:32:50 |
Impact of atherosclerosis on the reliability of LiDCOrapid
The relation between atherosclerosis and LiDCOrapid
Impact of atherosclerosis on the reliability of LiDCOrapid
The relation between atherosclerosis and LiDCOrapid
Japan |
Patients who will undergo cardiovascular surgery under median sternotomy
Operative medicine |
Others
NO
To uncover the influence of atherosclerosis on the reliability of LiDCOrapid
Efficacy
cardiac output, cardiac index
Observational
20 | years-old | <= |
Not applicable |
Male and Female
Patients who receive arterial line and pulmonary artery catheter. Patients who undertake cardio ankle vascular index (CAVI) and ankle brachial index (ABI) measurement.
Emergent, hemodialysis, severe AR, IABP, severe arrhythmia, ABI< 0.9
48
1st name | |
Middle name | |
Last name | Hirata Naoyuki |
Sapporo Medical University
Department of Anesthesiology
south1 west16 chuo-ku sapporo
011-611-2111
naohirata@mac.com
1st name | |
Middle name | |
Last name | Yoshikawa Yusuke |
Sapporo Medical University
Department of Anesthesiology
south1 west16 chuo-ku sapporo
011-611-2111
yusuke_19_@hotmail.com
Department of Anesthesiology, Sapporo Medical University
none
Other
NO
2014 | Year | 05 | Month | 17 | Day |
Published
CO/CI measured by LiDCOrapid was significantly correlated with those by PAC. However, the result of Bland-Altman analysis was not acceptable.
Completed
2014 | Year | 03 | Month | 19 | Day |
2014 | Year | 05 | Month | 17 | Day |
we will measure CO/CI by PAC and LiDCOrapid during anesthesia
2014 | Year | 05 | Month | 17 | Day |