Unique ID issued by UMIN | UMIN000008320 |
---|---|
Receipt number | R000009796 |
Scientific Title | Comparison of intraocular pressure measurements on the bed by four portable tonometer including icare pro. |
Date of disclosure of the study information | 2012/07/15 |
Last modified on | 2013/02/25 11:05:56 |
Comparison of intraocular pressure measurements on the bed by four portable tonometer including icare pro.
Comparison of intraocular pressure on the supine position.
Comparison of intraocular pressure measurements on the bed by four portable tonometer including icare pro.
Comparison of intraocular pressure on the supine position.
Japan |
Healthy volunteers
Ophthalmology |
Others
NO
To evaluate intraocular pressure on the bed among four portable tonometers and thier agreements
Efficacy
Intraocular pressure
Observational
20 | years-old | <= |
80 | years-old | >= |
Male and Female
1 Healty volunteers
2 Subjects who sign an informed consent form
1 Subjects with any ocular diseases.
2 Subjects with previous any ocular surgery.
120
1st name | |
Middle name | |
Last name | Shunsuke Nakakura |
Saneikai Tsukazaki Hospital
Ophthalmolgy
Himeji City Aboshi Waku 68-1
1st name | |
Middle name | |
Last name | Shunsuke Nakakura |
Saneikai Tsukazaki Hospital
Ophthalmolgy
Himeji City Aboshi Waku 68-1
+81-79-272-8555
shunsukenakakura@yahoo.co.jp
Saneikai Tsukazaki Hospital
None
Self funding
NO
2012 | Year | 07 | Month | 15 | Day |
Unpublished
The repeatability of Icare PRO was slightly lower in the supine position than in the sitting position. Although Icare PRO underestimated IOP values in eyes with higher IOP when compared with Tonopen XL and Kowa HAT in both positions, we observed good inter-device agreement between Icare PRO and both Tonopen XL and Kowa HAT.
Completed
2012 | Year | 07 | Month | 03 | Day |
2012 | Year | 07 | Month | 01 | Day |
2012 | Year | 10 | Month | 01 | Day |
2012 | Year | 12 | Month | 01 | Day |
2012 | Year | 12 | Month | 01 | Day |
2013 | Year | 01 | Month | 31 | Day |
To evaluate intraocular pressure by icare pro, TonopenXL, Diaton tonometer and handled applanation tonometer
2012 | Year | 07 | Month | 02 | Day |
2013 | Year | 02 | Month | 25 | Day |
Value
https://center6.umin.ac.jp/cgi-open-bin/ctr_e/ctr_view.cgi?recptno=R000009796