Unique ID issued by UMIN | UMIN000005287 |
---|---|
Receipt number | R000006278 |
Scientific Title | Comparative study of human choroidal structure by different OCT machines |
Date of disclosure of the study information | 2011/03/22 |
Last modified on | 2012/02/08 21:24:06 |
Comparative study of human choroidal structure by different OCT machines
Choroidal structure by OCT
Comparative study of human choroidal structure by different OCT machines
Choroidal structure by OCT
Japan |
Normal eye
Ophthalmology |
Others
NO
To see the difference of choroidal sturucure examined by different OCT machines
Others
Choroidal thickness measured by 3 different OCT mahcines
Confirmatory
Pragmatic
Not applicable
Choroidal thickness measured by three different OCTs
Difference of choroidal thickness based upon gender, age and other factors
Observational
18 | years-old | <= |
40 | years-old | >= |
Male and Female
normal eye
Highly myopic eyes -5D or greater
Those of which fundus cannot be observed.
Those with apparent ocular diseases such as glaucoma, diabetic retinopathy, macular degenerations and others.
40
1st name | |
Middle name | |
Last name | Taiji Sakamoto |
Kagoshima University
Ophthalmology
8-35-1 Sakruagaoka, Kagoshima 890-8520
+81-99-275-5402
1st name | |
Middle name | |
Last name | Toshifumi Yamashita |
Kagoshima University
Ophthalmology
8-35-1 Sakruagaoka, Kagoshima 890-8520
+81-99-275-5402
http://www.kufm.kagoshima-u.ac.jp/~op/gairai/RCstructurestudy.html
yyamatos@m.kufm.kagoshima-u.ac.jp
Ophthalmology, Kagoshima Univ.
Ophthalmology, Kagoshima Univ.
Japan
NO
鹿児島大学病院(鹿児島県)
2011 | Year | 03 | Month | 22 | Day |
http://www.kufm.kagoshima-u.ac.jp/~op/gairai/RCstructurestudy.html
Published
http://www.iovs.org/content/early/2012/01/10/iovs.11-8836.long
Results:Forty-three subjects (mean age 30.5 years) were enrolled. Of 43 the eyes, SCT of 39 eyes (90.7%) could be measured using each machine. Intra-class correlation coefficients (95% confident interval) were 0.976 (0.954-9.987), 0.958 (0.919-0.978) and 0.939 (0.895-0.971) with Spectralis, Cirrus, and Topcon respectively. Inter-rater correlation coefficients (95% confident interval) were 0.944 (0.893 to 0.971), 0.956 (0.831 to 0.983) and 0.924 (0.825 to 0.964) with Spectralis, Cirrus, and Topcon respectively. The average SCT was 272.6, 272.8, and 269.2 µm with Spectralis, Cirrus, and Topcon, respectively. Inter-machine correlation coefficient was significantly high among the machines (P<0.001, Spearman); 0.97 (Spectralis-Cirrus), 0.96 (Cirrus-Topcon) and 0.98 (Topcon-Cirrus). Bland-Altman plot analysis showed no typical trend among the machines.Conclusions:SCT measurements obtained with 3 different SD-OCTs were highly correlated and could be used interchangeably.
Completed
2011 | Year | 02 | Month | 01 | Day |
2011 | Year | 03 | Month | 01 | Day |
2011 | Year | 04 | Month | 01 | Day |
2011 | Year | 04 | Month | 01 | Day |
2011 | Year | 04 | Month | 01 | Day |
2011 | Year | 04 | Month | 01 | Day |
Comparison of results of choroidal thickness using several different OCT machines.
Publishsed as follows:
Repeatability and reproducibilityof subfoveal choroidal thickness in normal eyes of Japanese using different SD-OCT devices.
Yamashita T, Yamashita T, Shirasawa M, Arimura N, Terasaki H, Sakamoto T.
Invest Ophthalmol Vis Sci. 2012 Jan 12. [Epub ahead of print]
2011 | Year | 03 | Month | 22 | Day |
2012 | Year | 02 | Month | 08 | Day |
Value
https://center6.umin.ac.jp/cgi-open-bin/ctr_e/ctr_view.cgi?recptno=R000006278