Unique ID issued by UMIN | UMIN000002780 |
---|---|
Receipt number | R000003382 |
Scientific Title | Validation of the extravascular lung water by single transpulmonary thermodilution |
Date of disclosure of the study information | 2009/11/19 |
Last modified on | 2010/09/09 08:53:30 |
Validation of the extravascular lung water by single transpulmonary thermodilution
Validation of the extravascular lung water by the PiCCO system
Validation of the extravascular lung water by single transpulmonary thermodilution
Validation of the extravascular lung water by the PiCCO system
Japan |
Pulmonary edema
Pneumology | Emergency medicine | Intensive care medicine |
Others
NO
Validate the accuracy of extravascular lung water (EVLW) measured by the PiCCO system
Efficacy
Investigate the relationship between the EVLW measured by the PiCCO system and postmortem lung weight.
Observational
Not applicable |
Not applicable |
Male and Female
We studied consecutive autopsy cases whose EVLW were measured by the PiCCO system just prior to death
We chose 48 hours as a cut-off point for inclusion in the study because the postmortem lung weight is known to change little in the early postmortem period (4.5-72 hours)
30
1st name | |
Middle name | |
Last name | Takashi Tagami |
Aizu chuo hospital
Department of Emergency & Critical Care Medicine
1-1 Tsuruga, Aizuwakamatsu, Fukushima, Japan
1st name | |
Middle name | |
Last name |
Aizu chuo hospital
1-1 Tsuruga, Aizuwakamatsu, Fukushima, Japan
tagami.takashi@mbp.nifty.com
Aizu chuo hospital
Aizu chuo hospital
Self funding
NO
2009 | Year | 11 | Month | 19 | Day |
http://ccforum.com/content/14/5/R162
Published
http://ccforum.com/content/14/5/R162
A definite correlation exists between EVLW measured by the single-indicator transpulmonary thermodilution technique and post-mortem lung weight in humans. The normal EVLW value is approximately 7.4 +/- 3.3 mL/kg.
Completed
2004 | Year | 07 | Month | 29 | Day |
2004 | Year | 07 | Month | 01 | Day |
2009 | Year | 12 | Month | 01 | Day |
We studied the relationship between the EVLW measured by the PiCCO system and postmortem lung weight. Linear regression line equation was calculated if the relationship was statistically significant.
2009 | Year | 11 | Month | 19 | Day |
2010 | Year | 09 | Month | 09 | Day |
Value